A Python implementation of Variable-Rate Selective Excitation (VERSE) for MRI pulse design. Upload any RF pulse waveform and compress it to the shortest possible duration while respecting hardware limits (RF amplitude, gradient strength, slew rate). Based on the algorithm by Hargreaves et al. (2004).
Try it in your browser — no install needed. Run the built-in example (the same TB=10 sinc shown below) to see VERSE in action, or upload your own pulse and download the compressed version.
In MRI, a selective RF pulse is played together with a gradient to excite a specific slab of tissue. The excitation profile (which tissue gets excited, and by how much) depends on the integral of the RF waveform over time. This means you can make the pulse shorter — as long as you scale up the RF amplitude and gradient to compensate, the excitation profile stays the same.
The tradeoff: compressing the pulse increases the RF amplitude, the gradient strength, and how fast the gradient changes (slew rate). All three have hardware safety limits. So the question becomes: how short can we make the pulse before we hit a limit?
The VERSE approach: compress the pulse non-uniformly. At moments where the RF is weak (e.g., zero-crossings of a sinc), there's room to compress aggressively. At the peaks, the RF amplitude is already high, so less compression is possible. The optimal pulse hits at least one safety limit at every moment in time.
A TB=10 windowed sinc pulse compressed from 2.9 ms to 0.72 ms (4x) while respecting all hardware limits.
A standard selective RF pulse is played with a constant gradient. VERSE introduces a time-dilation function that "speeds up" time at each point. To preserve the excitation profile, the RF and gradient must scale accordingly:
The algorithm finds the maximum
| Constraint | Limit | What it protects |
|---|---|---|
| RF amplitude | RF amplifier | |
| Gradient amplitude | Gradient coils | |
| Gradient slew rate | Gradient amplifier / patient safety |
The gradient naturally ramps up from zero and back down -- no separate ramp handling needed. The slew rate constraint creates these ramps automatically.
Final VERSE pulse with all three constraints shown. The slew rate stays within limits throughout.
See 20260329 VERSE Outline.md for a detailed walkthrough of the algorithm and the math behind each step.
pip install -r requirements.txt
python verse_optimizer.pyThis runs the default example (TB=10 windowed sinc, 60° flip angle) and produces:
- Diagnostic plots at each algorithm step
- A comparison plot (original vs VERSE)
- An output file
verse_pulse_output.txtwith the compressed pulse
Three premade pulses are included. Edit the selection block near the top of verse_optimizer.py:
B1, G_orig, dt, pulse_name = make_sinc_pulse(TB=10, T_pulse=2.9e-3, flip_angle=60)
# B1, G_orig, dt, pulse_name = make_sinc_pulse(TB=5, T_pulse=1.5e-3, flip_angle=60)
# B1, G_orig, dt, pulse_name = make_gaussian_pulse(T_pulse=2.0e-3, flip_angle=60)
# B1, G_orig, dt, pulse_name = make_block_pulse(T_pulse=0.5e-3, flip_angle=90)Adjust these at the top of the script to match your scanner:
B1_max = 15.0 # Maximum RF amplitude, uT
G_max = 40.0 # Maximum gradient amplitude, mT/m
S_max = 150e3 # Maximum gradient slew rate, mT/m/sThe output file (verse_pulse_output.txt) contains tab-separated columns:
| Column | Unit | Description |
|---|---|---|
time_us |
us | Time axis |
B1_uT |
uT | RF amplitude |
G_mTm |
mT/m | Gradient amplitude |
Load in MATLAB: data = load('verse_pulse_output.txt');
Load in Python: data = np.loadtxt('verse_pulse_output.txt')
The code is written in NumPy array style (no classes, no Python-specific idioms) for straightforward translation to MATLAB. Key mappings:
| Python/NumPy | MATLAB |
|---|---|
np.sinc(x) |
sinc(x) |
np.hamming(N) |
hamming(N) |
np.interp(xq, x, y) |
interp1(x, y, xq) |
np.diff(x) |
diff(x) |
np.minimum(a, b) |
min(a, b) |
a[0] (0-indexed) |
a(1) (1-indexed) |